Shopping Cart

News — Manual Sampling

Why should I use a Cannula Stopper for Manual Sampling?

Posted by Pam Bialiy on

Why should I use a Cannula Stopper for Manual Sampling?

The following article has been authored by John Heaney. While automation is prevalent with dissolution testing, manual testing can and still should be done in certain circumstances.  This can be when automation is too expensive, when the method is being developed and a certain amount of flexibility is needed, or when sample times are too short to be completed by automated systems.  Manual sampling does offer significant flexibility but it also has some factors that need to be considered to ensure consistency. Sampling height is clearly defined in USP <711> as half-way between the top of the apparatus, be it...

Read more →


Hydrodynamic Effects of Resident vs Manual Probes/Cannula

Posted by Pam Bialiy on

Hydrodynamic Effects of Resident vs Manual Probes/Cannula

The following article has been authored by John Heaney. Does the presence of a resident sampling probe affect the hydrodynamics of a dissolution vessel?  Yes, yes it does.  The more important questions are how strong is the change in the hydrodynamics and is it strong enough to affect testing? USP <1092>, an advisory chapter addresses this directly with:  Sampling probes may or may not remain in the vessel throughout the entire run. Sampling probes or fiber-optic probes can disturb the hydrodynamics of the vessel; therefore, adequate validation should be performed to ensure that the probes are not causing a significant...

Read more →

Categories